AC current: 200A, 600A; 1.0+3 at 600A 50/60Hz
AC voltage: 200A, 600A; 1.0+2 (50/60Hz)
Resistance: 200Ω; 1.2+4, Beeps at below 30Ω (continuity check)
Method of detection: Mean value
Display: LCD(Digital display:4000 counts)
Response time: Approx. 2 seconds
Range switching: Manual-range(on AC current range)/Auto-range(on AC voltage range, resistance range)
Data hold: On all range
Peak hold: On AC current range
Operating temperature and humidity: 0-40;, 85% RH or less (no condensation)
Effect of external magnetic field :1A or less at 400A/m
Effect of conductor position: (2.0% rdg + 3dgt) or less
Safety standard Conforms: EN61010-1, EN61010-2-031, EN61010-2-032
Circuit voltage: 1000Vrms or less
Withstanding voltage: 6300V AC for 5 sec
Power supply: R6P(SUM-3) 2 or LR6 2
Battery life: Approx. 150 hours (continuous)
Consumed current: Approx. 5mA
Sleep function: Automatically powered down in about 10 minutes after the last switch operation
Diameter of measurable conductor: 54.5mm at maximum
Dimensions: Approx. 105(W) 247(H) 49(D)mm
Weight: Approx. 470g
Accessories: User's manual, batteries, carrying case (93034), Output plug (98012), Test Lead (98072)