Pressure range with ball-valve:
0 … 40 bar / 0 … 580 psia
Up to 120 °C (248 °F) and 40 bar
Adjustable length: 35 … 157/379 mm (1.37 … 6.2 /14.9 in)
Temperature measurement range: -40 … +180 °C (-40 … 356 °F)
Installation
Fitting body: R1/2" ISO
Fitting body: NPT 1/2"
Ball-valve set BALLVALVE-1
Sampling cell: DMT242SC2
Water activity
Measurement range: 0 … 1 aw
Response time (90 %) at +20 °C in still oil (with stainless steel filter): 10 min
Sensor: HUMICAP 180L2
Accuracy (including non-linearity, hysteresis, and repeatability):
0 … 0.9: ±0.02
0.9 … 1.0: ±0.03
Temperature
Measurement range: -40 … +180 °C (-40 … +356 °F)
Accuracy at +20 °C (+68 °F)
Display: LCD with backlight, graphical trend display of any parameter
Menu languages: English, Chinese, Finnish, French, German, Japanese, Russian, Spanish, Swedish
Optional data logger with real-time clock
Logged parameters: Max. four with trend/min./max. values
Logging interval: 10 sec (fixed)