Platform specifications
Temperature Range:
-100°C to +200°C (LN2)
-60°C to +200°C (LCO2 - 900 PSI)
-40°C to +200°C (LCO2 - 300 PSI)
Temperature Transition Rate:
Cooling: 19°C-40°C/minute Typ
Heating: 13°C-29°C/minute Typ.
Workspace Dimensions: 22.0”W x 18.0”D (55.9cm x 45.7cm)
Workspace Surface Area: 396 in2 (2,555 cm2)
External Dimensions: 3.2”H x 23.0”W x 23.5”D (8.1cm x 58.4cm x 60.0cm)
Mounting Style: Benchtop / Optional: Shelf Rack Mounting
Coolant Type: LN2 (LCO2 optional, reduced low end temp)
Power Requirements (can be configured for CE): 200-240V / 50-60HZ / 1PH / 32A
Options:
• Delivery Hoses: Vacuum Jacketed; others
• Independent Failsafe Module (IFM) (on Controller, see Page 2) with Redundant Solenoid Valves (RSV) Shuts off coolant to prevent runaway cold condition
• Covers, Adapter Plates, and Additional Hole Patterns
• Dry Nitrogen Purge for Moisture Control
• Dewar Accessories: Distribution Manifolds, Pressure Regulators
Controller Specifications
Temperature Measurement:
Range: -210°C to 680°C
Accuracy: ±1.0°C
Resolution: 0.1°C full scale
User Interface: 5.7” color touch-screen with temperature graphing and charting
Control Safety: Independent Fail-Safe Module (IFM) (optional) high and low temperature limits
Compliance: CE / RoHS / Designed to meet UL 61010
Diagnostics: Controller, chiller, & blower runtime hours; Valve activation counts; Controller enclosure temperature; System performance log
Operating Environment:Temperature: 10°C to 50°C; Humidity: 10% to 50%
Temperature Inputs: RTD (500 Ohm/2Wire) (100 Ohm/3 Wire); Thermocouple (type K)
Control Algorithms: Primary loop PID, Dual loop, settable DUT control
Communication Interfaces: Ethernet 10/100 / Telnet / HTML Web Server / USB-2.0 ; Optional: RS232 and IEEE-488 GPIB
Power Requirements: Voltage: 100 to 250 VAC; Frequency: 50/60 Hz; Current: up to 30A (application based)
Physical Dimensions: (TS1-4) 8.5”H x 6”W x 11.6”D (21.6cm x 15.2cm x 29.5cm); (TS5-8) 8.5”H x 8.0”W x 11.6”D (21.6cm x 20.3cm x 29.5cm)
Program Compatibility: Supports Sigma C, CC3, CC3.5, C4, and C5 functionality & command set
TS5-TS8 Options: Purge Set Point, Flow Meter, ON/OFF Solenoid Valve