Detector
Energy compensated Si-Diode
Button Functions
• change display (dose/dose rate)
• switch ON/OFF
• chirp ON/OFF
• reset integrated dose
• change alarm thresholds
• activate battery test
Measurement Range
Dose: 1 uSv - 9.99 Sv or 0.1 mrem - 999 rem
Energy Response: Hp(10), 60 keV - 3 MeV, better than + or - 25%, up to 6 MeV, better than + or - 35%
Dose Rate: 5 uSv/h - 3 Sv/h or 0.5 mrem/h - 300 rem/h
Dose Rate Linearity: Better than + or - 15% up to 3 Sv/h (300 rem/h)
Calibration: Better than + or -5% (Cs-137, 662 keV at 2 mSv/h), Hp(10)
Audible Alarms
Seven separate alarms, sound level typically better than 85 dBA at 30 cm
• integrated dose
• dose rate
• dose overflow
• dose rate overflow at 3 Sv/h or 300 rem/h
• low battery 1 and 2
• defect
Alarm Thresholds: Six preset values each for integrated dose and dose rate-push button selection
Connectors: Infrared communication via bottom of the dosimeter
Power Requirements: One triple A alkaline cell, typical life is 1800 hours in background
Temperature Range
-20 - + 50° C operational, humidity up to 90% RH, non-condensed
-20 - + 70° C storing
Weight: 80 g (including battery)
Size: 78 x 67 x 22 mm