Simulation Type
Lead I: 70%
Lead II: 100%
Lead III: 30%
Lead V1: 24%
Lead V2: 48%
Lead V3: 100%
Lead V4: 120%
Lead V5: 112%
Lead V6: 80%
ECG Performance Tests
Square Waves: 2Hz, 0.125Hz
Triangle Wave: 2Hz
Pulse: 60bpm or 240bpm
Sine Waves: 0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100Hz
Performance Amplitude: 0.5 to 5.0mV in 0.5mV steps
Arrhythmias
Ventricular:
Asystole
Bigeminy
Trigeminy
Ventricular fibrillation (coarse)
Ventricular fibrillation (fine)
Ventricular tachycardia
Supraventricular:
Atrial fibrillation (coarse)
Atrial fibrillation (fine)
Atrial flutter
Atrial tachycardia
Missing beat
Nodal rhythm
Paroxysmal atrial tachycardia
Sinus arrhythmia
Supraventricular tachycardia
Atrial Conduction:
First degree AV block
Left bundle branch block
Right bundle branch block
Second degree AV block - Mobitz I
Second degree AV block - Mobitz II
Third degree AV block
Premature:
Premature atrial contraction
Premature nodal contraction
Premature left ventricle contraction
Premature left ventricle contraction - early
Premature left ventricle contraction - R on T
Premature right ventricle contraction
Premature right ventricle contraction - early
Premature right ventricle contraction - R on T
Premature ventricular contraction - 6 / min
Premature ventricular contraction - 12 / min
Premature ventricular contraction - 24 / min
Premature ventricular contraction - frequent multifocal
Invasive Blood Pressure
Channels: 4 channels
Static Pressure: -10, -5, 0, 20, 40, 50, 60, 80, 100, 150, 160, 200, 240, 320, 400mmHg
Dynamic Simulation:
Arterial [ART]: 120/80
Radial Artery [RA]: 120/80
Left Ventricle [LV]: 120/00
Right Ventricle [RV]: 25/00
Right Atrium (central venous) [CVP]: 15/10
Pulmonary Artery [PA]: 25/10
Pulmonary Artery Wedge [PAW]: 10/2
Left Atrium [LA]: 14/4
Accuracy: ± 1mmHg
Excitation Voltage: 2 to 16V
Impedance: 350Ω Nominal
Simulated Sensitivity: 5μV/V/mmHg or 40μV/V/mmHg (user selectable)
Connector: Mini DIN style
Respiration
Rates: Selectable values of 0, 5, 10, 15, 30, 40, 60, 80, 120, 180 brpm
Lead Selection: Lead I and II, user selectable
Base Resistances: 500, 1000, 1500 & 2000Ω
Accuracy: ±10%
Resistance Variations: 0.2, 0.5, 1.0, 3.0Ω
Apnoea Simulation: Manual ON / OFF
ECG
Normal Sinus Rhythm:
Rates: 30, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 BPM
Rate Accuracy: ±1BPM or 1%
Amplitude (Lead II): 0.05 to 0.45mV (0.05mV steps), 0.5 to 5.5mV (0.5mV steps)
Amplitude Accuracy: ± 2%
ST Segments: 18 selectable values, 8 elevated and 8 depressed -0.8 to +0.8mV in 0.1mV steps and +0.05mV and -0.05mV on Lead II
Neonatal Mode: ECG R wave width is reduced to 40ms
Pacemaker
Simulated Rhythms:
Asynchronous at 75 BPM
Demand with frequent sinus beat
Demand with occasional sinus beat
A-V sequential
Non-capture
Non-function
Pulse Amplitude: 1.0, 2.0, 5.0, 10.0mV
Accuracy: ± 5%
Width: 0.1, 0.2, 0.5, 1.0, 2.0ms
R-Wave Detector Test: 30, 60, 80, 120, 200, 250 BPM
Haver-Triangle Width: 8, 10, 12, 20, 40, 60, 80, 100, 120, 140, 160, 180, 200ms
Performance Amplitude: 0.5 to 5.0mV in 0.5mV steps
Fetal Simulation
Maternal Heart Rate: 80 BPM (Fixed)
Fetal Heart Rate (Selectable): 60, 90, 120, 140, 150, 210, 240 BPM
Fetal Heart Rate (IUP): 140 BPM at beginning, then varying with pressure
Intrauterine-Pressure Waveforms: Early deceleration, late deceleration, and uniform acceleration
Simulation Period: Manual, or 2, 3, or 5 minutes
Temperature
Simulation: YSI 400 / 700A / 700B Static
Temperature Unit: °C or °F, user selectable
Range: 0.0, 24.0, 37.0, and 40.0°C; 32.0, 75.2, 98.6, 104.0°F
Accuracy: ± 0.1°C / °F
Connector: Mini DIN style
Mains Power/Battery Info
Battery: 3.7V 3900mAh 14.4WH
Power Supply: 5V 1A USB micro-B
Input Voltage: 100-240V ~ 50/60Hz 0.18A max.
Charge Time (new battery): Up to 3-4 hours
Battery Life: Up to 6 hours, ~4 hours
Weight: 0.70kg / 1.5lbs
Dimensions: 180 x 150 x 55mm, 7.1 x 5.9 x 2.2"
Cardiac Output
Catheter Type: Baxter Edwards, 93a-131-7f
Calibration Coefficient: 0.542 (0 °C injectate), 0.595 (24 °C injectate)
Blood Temperature: 37 °C (98.6 °F) ± 2 %
Injectate Volume: 10 cc
Injectate Temperature: 0 °C or 24 °C ± 2 % value
Cardiac Output: 2.5 l/min, 5 l/min, 10 l/min ± 5 %
Connector: Mini DIN style