Energy range of the measures of photon radiation: 60 keV - 3 MeV
Ambient dose equivalent rate (ADER) of photon radiation measurement range: 0.1 - 30 μSv/h (for energy 662 keV)
Accuracy of photon radiation ADER measurement: ±30 %
False alarm rate: 1/1000
Minimum detectable amounts of radioactive materials:
Control zone (L x H) 1.5 x 3 m:
Conditions: with 0.5 probability and at least 0.95 confidence probability, at gamma radiation background 0.1 μSv/h, at source speed 5 km/h
133Ba: 40 kBq
137Cs: 50 kBq
60Co: 24 kBq
Control zone (L x H) 3 x 3 m:
Conditions: with 0.5 probability and at least 0.95 confidence probability, at gamma radiation background 0.1 μSv/h, at source speed 15 km/h
133Ba: 165 kBq
137Cs: 200 kBq
60Co: 95 kBq
Control zone (L x H) 3 x 3.5 m:
Conditions: with 0.5 probability and at least 0.95 confidence probability, at gamma radiation background 0.25 μSv/h, when source is shielded with 1 mm of steel
At source speed 10 km/h:
133Ba: 340 kBq
137Cs: 410 kBq
60Co: 210 kBq
In long-dwell scan mode:
133Ba: 60 kBq
137Cs: 70 kBq
60Co: 35 kBq
Control zone (L x H) 6 x 3 m:
Conditions: with 0.5 probability and at least 0.95 confidence probability, at gamma radiation background 0.1 μSv/h
At source speed 15 km/h:
133Ba: 315 kBq
137Cs: 390 kBq
60Co: 190 kBq
In long-dwell scan mode:
133Ba: 53 kBq
137Cs: 67 kBq
60Co: 32 kBq
Number of detection units: 2
PC communication: Ethernet, WiFi
Ingress protection: IP65
Mean recovery time: no more than 60 min
Mean service lifetime: 15 years
Mean time between failures, no less than: 20 000 h
Atmospheric pressure: from 84 to 106.7 kPa
Operating temperature:
Standard: -30 up to +50 °C
With heating module: -50 up to +50 °C
Humidity: 98 % at +40 °С and lower temperature
Power source: AC mains, built-in accumulator
Supply voltage: 90 - 132 V or 180 - 264 V, 47 - 63 Hz
Power consumption:
Without video surveillance and heating: no more than 80 V⋅A
With video surveillance and heating: no more than 940 V⋅A
Battery Lifetime (on accumulator): 3 h