Measuring principle: Near-infrared 90° scattered light measurement method

Indication: LCD4-digit

Measuring range:

Turbidity: 0~500°C

SS:0~500mg/L

*Specification: 0~1000

Resolution:

Range of 0.1 /0~199.9

Range of 1 /200~500

Full scale For 1000: 1 For

Full scale 20: 0.01

Reproducibility: ± 2%FS or less (under certain conditions)

SS converter:

SS conversion factor can be set

(x = turbidity, y=SS and y=a+bx)

Zero calibration:

1st standard: pure water

2nd standard: simple zero calibration plate

Span Calibration:

1: Set the SS conversion factor after calibration with formazine turbidity standard solution

2: Direct SS calibration 1 or 2 point calibration of the SS turbidity of the sample water after manual analysis

Self-diagnostic function: Battery voltage, calibration abnormality, detector abnormality

Ambient temperature range: 0~40℃

Power: AAA alkaline batteries (LR03×3 batteries) with auto power-off function