Measurement channels: 2
Display: LCD, LED backlight
Rate 1/s: 10
Connectable gauges with display range
CDG (A/D)Torr: 1×10^-3×F.S. …1×F.S
PCG (mbar): 5×10^-4…1500
PSG (mbar): 5×10^-4…1000
MPG (mbar): 1×10^-9…1000
PEG (mbar): 1×10^-9…1×10^-2
MAG (mbar): 1×10^-9…1×10^-2
BCG (mbar): 5×10^-10…1500
BPG (mbar): 5×10^-10…1000
HPG (mbar): 2×10^-6…1000
Measurement unit (selectable): mbar, Torr, Pa, hPa, Micron, V
Setpoints
Setpoint relays: 4
Channel assignment: 1 or 2
Adjustment range: sensor dependent
Hysteresis: adjustable
Relay contact: floating changeover contact
Setpoints Contact rating V (ac) • A: 30 / 1
Setpoints Contact rating V (dc) • A: 60 / 0.5
Connector: D-Sub, 9-pin, male
Analog output
Range: 0 ... 10.3 V, sensor analog output signal
Analog output: 2
Programmable analog output: 1
Connector: D-Sub, 9-pin, male
Interface (digital) Connector: USB slave, master and Ethernet, USB Typ A (stick), USB Type B, FCC68/RJ45
Power Supply: 100 ... 240 V (ac)
Frequency: 50 ... 60Hz
Consumption: ≤65W
Operation temperature (ambiance): +5 ... +50°C