Mainframe
Microprocessor controlled touch panel: 7”, capacitive
Optical Ethernet Interface for remote control of the generator: optional
Interface for saving reports: USB
External trigger input/output: Switch / 10 V
Connector for external safety interlock loop: 24 V =
External red and green warning lamps: 24 V=, 40 mA
Mains power: 90 V - 264 V, 50/60 Hz
Dimensions of desk top case (W × H × D): 450 × 330 × 500 mm³
Weight: 25 kg
Internal 1-phased de-/coupling network (CDN)
Coupling-/decoupling network for power supply lines: L1, N, PE
Nominal voltage, nominal current: 300 V, 16 A ac/dc ≈ / 300 V, 16 A =
Coupling impedance (depending on the generator): 33 nF / 18 µF / 9 µF + 10 Ω
RINGWAVE (option) acc. to IEC 61000-4-12: 2017
Impulse output voltage, adjustable: 0.2 - 7.0 kV ± 10 %
Output wave form acc. to IEC 61000-4-12: 0.5 µs / 100 kHz
Oscillation frequency: 100 kHz ± 10%
Voltage rise time to the first peak (open): 0.5 µs ± 30%
Current rise time to the first peak (short): ≤ 1 µs
Decaying voltage:
Ratio of Pk2 to Pk1: 0.4 < ratio < 1.1
Ratio of Pk3 to Pk2: 0.4 < ratio < 0.8
Ratio of Pk4 to Pk3: 0.4 < ratio < 0.8
Polarity of output voltage: pos / neg, selectable
Maximum stored energy: 12 J
Repetition rate, max.: 60 transients per minute
High-voltage output (HV-OUT), impedance selectable:
PFN 1: Series resistor 12 Ω (10+2 Ω)
PFN 2: Series resistor 30 Ω (10+20 Ω)
Monitor output for pulse output voltage: Ratio = 1000:1 ± 5%
Mains synchronous trigger: Phase shifting, digitally selectable 0 - 360° ±5°