HAIDA HD-E702-225 Constant Temperature and Humidity Test Chamber (-40℃~150℃, 20~98% RH)
Internal Dimension (W × H × D): 500 × 750 × 600 mm
External Dimension (Stainless Steel Type) (W × H × D): 1100 × 1670 × 1200 mm
External Dimension (Paint Type) (W × H × D): 800 × 1670 × 1200 mm
Total Power (kW): 9.5 / 9.5 / 11.6
Maximum Current (A): 25 / 32 / 40
Voltage: Three-phase 380 V
Performance:
Temperature & Humidity Adjustment Method: Balanced Temperature and Humidity Control (BTHC), PID intelligent adjustment
Temperature Range: -40℃~150℃
Temperature Accuracy: 0.01 °C
Temperature Tolerance: ≤ ±1.0 °C or ±2.0 °C
Temperature Fluctuations: ≤ ±0.5 °C (without load and temperature stable)
Temperature Uniformity: ≤1.5 °C (without loading), 50 mm ≤ Sensor
Humidity Range: 20% ~ 98% RH
Humidity Accuracy: 0.1% RH
Humidity Tolerance:
≥75% RH: ≤ ±3% RH
≤75% RH: ≤ ±6% RH
Humidity Fluctuations: ≤ ±2.5% RH
Heating Rate: 3 °C/min average
Cooling Rate: 1 °C/min average
Regulator:
Cooling Method: Single-stage compression, two-stage compression
Refrigerator: Hermetically sealed France Tecumseh compressor or semi-hermetic BOCK compressor
Cooling Method: Air-cooled
Controller:
Operation Panel: Programmable LCD Touch Screen
Running Mode: Fixed Running, Pattern Running
Program Memory Capacity: 120 programmable groups, maximum 100 sections per group
Output: RS-232 (USB, optional)
Water Supply System:
Water Supply: Circulating Water
Water Tank Capacity:
Chamber internal capacity <800 L: 25 L × 1 pc
Chamber internal capacity >800 L: 25 L × 2 pcs
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts