Indicator: 2 to 18 ppm
Measuring Range:
+ 1 to 18 ppm, Number of Pump Strokes 2
+ 18 to 36 ppm, Number of Pump Strokes 1
Sampling Time: 1.5 minute per pump stroke
Colour Change: White → Yellow
Reaction Principle: 2CH3Br +I2O5+ H2S2O7 → Br 2, Br 2 +O-Tolidine → Yellow product
Coefficient of Variation: 10% (for 1 to 6 ppm), 5% (for 6 to 18 ppm)
Nr. of tests: 5
Pump: GV-100S/110S
Catalog
Manual
manual tube
Performance
| Measuring Range |
|
| Number of Pump Strokes |
|
| Correction Factor |
|
| Sampling Time |
1.5 minutes per pump stroke |
| Detecting Limit |
0.2 ppm (n=2) |
| Colour Change |
White → Yellow |
| Reaction Principle |
2CH3Br +I2O5+ H2S2O7 → Br 2 Br 2 +O-Tolidine → Yellow product |
| Coefficient of Variation |
10% (for 1 to 6 ppm), 5% (for 6 to 18 ppm) |
| Shelf Life |
2 Years |
| Corrections for temperature & humidity |
Unnecessary |
| Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
| Substance |
Concentration |
Interference |
Change colour by itself to |
| Hologenes, NOx |
|
+ |
Yellow |
| Saturated halogenated hydrocarbons |
|
+ |
Yellow |
Other substance measurable with this detector tube
| Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
| n-Butyl Bromide |
1.0 |
2 |
1 to 18 ppm |
| n-Butyl Bromide |
2.4 |
1 |
2.4 to 43.2 ppm |
| n-Propyl Bromide |
1.0 |
2 |
1 to 18 ppm |
| Chloro bromomethane |
0.7 |
2 |
0.7 to 12.6 ppm |
| Calibration gas generation |
Permeation tube method |
| TLV-TWA |
TLV-STEL |
Explosive range |
| 1 ppm (2014) |
- |
10 to 15% |